Thermo Mechanical Analysis
Typ
TMA Q400EM
Geräteklasse
Thermal Analyzers
Hersteller
TA Instruments
Anwendung
- Measurement of temperature-dependent sample expansion and derivation of the coefficient of thermal expansion (CTE) of polymeric materials
- Search for well adapted materials dependent to the application
- Characterization of aging effects due to temperature change, humidity/heat, salt fog or media removal
- Characterization as input variable for FEM simulations for lifetime models of microelectronic assemblies
Eigenschaften
Properties
- Maximum force: 2 N
- Minimum force: 0,001 N
- Force resolution: 0.001 N
- Path resolution: <0.5 nm
- Frequency range: 0.01 to 2 Hz*
- Temperature range: -80 to 1000 °C**
- Heating and cooling rate: 2 to 50 K/min***
- Temperature stability: 0.1 K
* For DTMA and MTMA measurements
** Usable temperature ranges -70-400 °C and 25-1000 °C
*** Depending on target temperature
MTMA measurements
- Modulated thermo-mechanical analysis
- Used for separate consideration of simultaneous effects of shrinkage and thermal expansion
- Via a temperature modulation (sinusoidal) the sample expansion can be decomposed into a reversible and non-reversible part by FFT
Load possibilities
- Expansion probe for the CTE measurement of most solid-state materials
- Macro expansion probe for CTE measurement of very soft materials
- Penetration probe for characterizing the melting behavior of coatings on their substrates
- 3-point bending probe for characterizing the deformation temperature of stiff materials

Beispiele
CTE measurement of metals or polymers

CTE measurement of stacked composite systems

Measurement of sample expansion - example of a stabilized material or fully cross-linked polymer

Example of high post-crosslinking (usually unfavorable for reliability)

Calculation of the coefficient of thermal expansion (CTE)